EasyManuals Logo

Agilent Technologies HP 83711B Calibration Guide

Agilent Technologies HP 83711B
275 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #43 background imageLoading...
Page #43 background image
In CW mode, phase noise degrades with frequency or phase modulation:
Harmonics:
<-55
dBc
at output levels
<+6
dBm
-50
-55
-60
Z-65
=-70
+-75
pilo
0 2
4 6
Carrier Frequency
(GHz)
a
10
Figure l-25. Typical 2nd Harmonic Levels Measured at Output Power of + 6
dBm
Non-Harmonic Spurious (13
kHz):
c-60
dEk (includes power supply and frequency synthesis
spurious).
Non-Harmonic
Spurious
(~3
khk):
<
-50
dBc
Sub-HarmonicsNone
Residual
FM:
At 1
GHz,
cl5
Hz
in 50
IIZ
to 15
kHz
bandwidth.
Residual
l?M
decreases
6
dB
per octave below
1
GHz.
8 12
Carrier Frequency (GHz)
16
20
Figure l-26.
Typical Residual FM Measured in 50
Hz
-
15
kHz
Bandwidth, CW Mode, with High
Stability Timebase (Option lE5)
HP
83731Bl32B
Specifications
l-25