DRAFT 1/19/17
Vacuum Products Division
INSTRUCTION MANUAL
IDP-7
Dry Scroll
Vacuum Pump
Manual No. X3807-90009
Revision A
January 2017

Questions and Answers

  • W
    william24Aug 4, 2025
    What to do if my Agilent Technologies Water Pump will not start?
    • H
      Hayden NewmanAug 4, 2025
      If your Agilent Technologies Water Pump won't start, several issues could be the cause. First, check the fuse and replace it if blown, identifying the cause of the overload. Make sure to check the line voltage and the voltage configuration of the pump. The motor thermal protector might be open, in which case you should allow the motor to cool and then identify the cause of the overload. Also, excessive voltage drop could be the problem, so check the size and length of the cable. Finally, there could be a defective motor; in this case, inspect it.
  • D
    dana70Aug 10, 2025
    Why is my Agilent Technologies IDP-7 making a hammering noise?
    • S
      schaefersteveAug 10, 2025
      If your Agilent Technologies Water Pump is making a hammering noise, it could be due to overheating or debris inside the pump. Check the ventilation to the pump and the ambient temperature to address overheating. Also, inspect the intake screen for debris, flush the pump, and if necessary, disassemble the pump to inspect it thoroughly.
  • S
    sanchezmitchellAug 18, 2025
    What to do if Agilent Technologies Water Pump has poor ultimate pressure?
    • L
      Lynn MccormickAug 18, 2025
      If your Agilent Technologies Water Pump exhibits poor ultimate pressure, several factors could be responsible. You may have a system leak, so locate and repair it. Water or solvent might be present in the pump; flush the pump with air or dry nitrogen, and consider installing a trap or filter. The gas ballast could be plugged; replace the breather vent. Worn-out seals can also cause this issue, requiring a pump rebuild. Lastly, poor conductance to the pump could be the problem, so replumb with shorter and/or larger diameter tubing.
  • T
    Tracey WagnerAug 26, 2025
    What to do if my Agilent Technologies IDP-7 Water Pump runs intermittently?
    • S
      Stephanie ParkerAug 26, 2025
      If your Agilent Technologies Water Pump runs intermittently, the motor thermal protector might be cycling open and closed. Allow the motor to cool down and identify the cause of the overload.

Summary

Warranty

Warranty Replacement and Adjustment

Procedures for making warranty claims and handling product returns.

Environmental Conditions, Compliance and Utility Requirements

Technical Information

Specifications

Detailed technical specifications of the IDP-7 vacuum pump.

Installation

Safety

Critical safety precautions to follow during installation.

Startup

Steps for starting the vacuum pump correctly.

Electrical Connections

Guidelines for connecting the pump to the power supply.

Mechanical Connections

Details on connecting the pump's inlet, exhaust, and gas ballast.

Operation

Startup Procedure

Steps for safely starting the vacuum pump.

Troubleshooting

Troubleshooting Chart

A table to diagnose and resolve common pump problems.

Maintenance

Tip Seal Replacement

Procedure and parts needed for replacing pump tip seals.