56 1260 Infinity II Binary LC System - Manual and Quick Guide
3
Quick Start Guide
Check Out the System
Table 3 Checkout method parameter settings Binary Pump (G7112B)
Parameter Value
Flow 0.8 mL/min
Solvent A Water
Solvent B ACN
Composition 40 %B (ACN)
Composition 60 %A (Water)
Stoptime 10 min
Pressure Limit 600 bar
Minimum Stroke Automatic
Timetable 2.5 min80 %B
Table 4 Checkout method parameter settings Vialsampler/Multisampler
(G7129A/G7167A)
Parameter Value
Injection 1 µL
Stoptime as pump
Draw speed 100 µL/min
Table 5 Checkout method parameter settings Multicolumn Thermostat (G7116A)
Parameter Value
Temperature (left) 40 °C
Temperature (right) combined
Stoptime as pump