15 Connect the finger-tight fittings of the inner tubing kit to the ports of the
diverter valve. The tubings must run into the ports of the diverter valve in
horizontal, parallel lines.
NOTE
The tubings must not be bent or twisted.
16 Connect the tubing with the label with the arrow (1) to the flow cell of the
detector. Insert the other tubing (2) into the waste container.
17 Re-install the tray(s) in the tray base.
18 Start the instrument.
19 Close the front cover.

