Replace the Valve to Needle Tubing
For bio-inert modules use bio-inert parts only!
When When contaminated, worn out or visibly damaged
Typically once every year
Tools required p/n Description
8710-1534 Wrench, open end, 4 mm
8710-0510 Wrench open 1/4 — 5/16 inch
Parts required p/n Description
Preparative-scale:
G1364-68605 Tubing Kit 40 – 100 mL/min, 1.2 mm ID
Analytical-scale and bio-inert:
G1364-68602 Tubing Kit 1 – 5 mL/min, 0.25 mm ID
Standard tubing kit for Analytical and Bio-inert Fraction
Collector
G1364-68601 Tubing Kit 1 mL/min, 0.15 mm ID (OPTIONAL)
Preparations Position the transport unit of the fraction collector in the Home position.
Remove all installed trays from the tray base.
Position the transport unit of the fraction collector in the Change Parts
position and turn off the instrument.
It might be more convenient to remove the needle from its carrier before
unscrewing the needle tubing.
WARNING
Personal injury
Risk of personal injury caused by the needle arm movement.
Keep fingers away from the needle area.

Other manuals for Agilent Technologies InfinityLab LC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies InfinityLab LC Series and is the answer not in the manual?

Agilent Technologies InfinityLab LC Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelInfinityLab LC Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals