Replace the Valve to Needle Tubing
For bio-inert modules use bio-inert parts only!
When • When contaminated, worn out or visibly damaged
• Typically once every year
Tools required p/n Description
8710-1534 Wrench, open end, 4 mm
8710-0510 Wrench open 1/4 — 5/16 inch
Parts required p/n Description
Preparative-scale:
G1364-68605 Tubing Kit 40 – 100 mL/min, 1.2 mm ID
Analytical-scale and bio-inert:
G1364-68602 Tubing Kit 1 – 5 mL/min, 0.25 mm ID
Standard tubing kit for Analytical and Bio-inert Fraction
Collector
G1364-68601 Tubing Kit 1 mL/min, 0.15 mm ID (OPTIONAL)
Preparations • Position the transport unit of the fraction collector in the Home position.
• Remove all installed trays from the tray base.
• Position the transport unit of the fraction collector in the Change Parts
position and turn off the instrument.
• It might be more convenient to remove the needle from its carrier before
unscrewing the needle tubing.
Personal injury
Risk of personal injury caused by the needle arm movement.
➔ Keep fingers away from the needle area.