4 Installation Note E4440-90318
Audio Input, Option 107
NOTE
1. The installation of this kit requires that some performance testing be performed in
order to assure the new option is functioning properly.
This installation note includes a list of required Performance tests. Completing the
performance tests does not guarantee the instrument meets all specifications.
2. The instrument end user must determine whether they need a full instrument
calibration following the installation of the kit. If a full calibration is required,
arrangements regarding the level of calibration must be made between the end user
and the calibration provider.
3. This option is licensed for one instrument model/serial number combination. The
license key will only install on the designated instrument.
4. Option 233, Measuring Receiver Personality must be present in the PSA and the
instrument must be in the Measuring Receiver mode, before the Audio Input will
function.
5. Option 122, Wide Bandwidth Digitizer in NOT compatible with Option 107.