Loading...
XGS-600 Gauge Controller
20
DRAFT 4/
15/14
IMG-100
Inverted
Magnetron
Most durable and fastest
responding HV gauge (cold
cathode)
General purpose-measures to 2 x 10-9 Torr
Small, reliable, extremely rugged - Will not burn out
Very fast response - < 20 milliseconds
Less accurate and repeatable than hot filament
Longer starting in high vacuum than a hot filament
Can be cleaned
FRG700/702 Combination full range gauge Active gauge - Pirani/Cold Cathode
FRG720/730 Combination full range gauge Active gauge - Pirani/Hot Filament
Ultra High 10-9 to 10-12
IMG-300
Inverted
Magnetron
Most durable and fastest
responding UHV gauge
(cold cathode)
Measures to 1 x 10
-11
Tor r
Bakeable with magnet to 250 °C
Bakeable without magnet to 450 °C
Reliable, rugged, will not burn out
Can be cleaned
UHV-24
(nude)
Most popular UHV gauge
(hot filament)
Measures to 1 x 10
-10
Tor r
Bakeable with magnet to 250 °C
Thoriated iridium filaments run cooler - less outgassing
Dual tungsten filaments run hotter -
More outgassing but measurements are more accurate
and repeatable
UHV-24p
(nude)
UHV gauge providing the
highest vacuum reading
Same as UHV-24 except it measures to 2 x 10
-11
Tor r
Table 2-2 Gauge Selection Chart (Continued)
Controller Gauge Card
Vacuum Regime
Convection
Hot Filament
Cold Cathode
Analog
Gauge
Type Recommendation Gauge Characteristics