XGS-600 Gauge Controller
DRAFT 4/15/14
29
SET-PTS (Set Points) Displays the setpoints activated. The setpoints are numbered on the
Set Point screen (page 29).
The setpoint number appears if all following conditions met:
The setpoint is assigned to the sensor,
The setpoint screen is set to AUTO for that setpoint,
And the setpoint has been activated.
Set Point Screen
To open this screen select SETPTS on the Main screen.
Figure 3-3 Set Point Screen
Set Point # This number corresponds to the sequence of eight rear panel Set
Point outputs. This is also the number that appears on the Main
screen if the setpoint is activated. These are fixed numbers and not
user changeable.
SETPT Select an available gauge in this list to assign to a Setpoint #. Only
two set points are allowed per sensor. If two setpoints are already
assigned to a sensor, that sensor does not appear in the list of
options for this field.
NONE is available as a choice. If selected, all pressures and delays
are set to zero and OFF.