XGS-600 Gauge Controller
68
DRAFT 4/
15/14
The XGS-600 sends ?FF as a response if the command or data is invalid, or if the command
length is incorrect. There is no response to a wrong address, or lack of termination
character.
Table B-1 Serial Command Set
Description Command
Response from
XGS-600 Notes
Read XGS contents #aa01 >hhhhhhhhhhhh
where
hh sensor board codes
are:
10 = Hot
Filament Ion Gauge
board (HFIG)
3A =
Inverted Magnetron
board (IMG)
40 =
convection board
(CNV) = Analog Board
4C
FE =
Empty slot
Lists
two-character codes
identifying the 6 board
slots in the system,
following board
installation sequence
from left to right (front
view).
Read pressure #aa02cn
>x.xxxE-xx
Read
software revision
#aa05 >hhhh,hhhh . . .
where hhhh represents
revision hh.hh
Revisions shown for
main board, then each
sensor board following
board installation
sequence from left to
right (front view).
System Reset #aa06 >
Read
Pressure Dump
#aa0F >x.xxxE-xx,x.xxxE-xx, .
. .
All sensor
readings in one
command. Readings
follow board
installation sequence
from left to right (front
view).
Set pressure units to
Torr
#aa1 0 >
Set pressure units to
mBar
#aa11 >

Related product manuals