XGS-600 Gauge Controller
DRAFT 4/15/14
69
Set pressure units to
Pascal
#aa12 >
Read pressure units #aa13 >xx
where:
xx = 00 is Torr
xx = 01 is mBar
xx = 02 is Pascal
Assign User Label to
Sensor
#aa14cnLABEL
where c sensor code
=I for ion gauges HFIG
and IMG
=T for CNVs + AUXs
n = Sensor Count
counting TCs or ion
gauges from left to right
from the front panel
view.
LABEL= user-defined
1-5 characters.
> Label can be
any 5 or fewer
characters (AHZ, 0-9,
or space), except for
IMGxx, HFIGx,
CNVxxx and AUXxxx,
(where x is any
character) which are
reserved.
Read User Label of
Sensor
#aa15cn >LABEL
where LABEL is
user-defined 1-5
characters.
If no User Label is
assigned, defaults to
system-assigned Sensor
ID.
Setup Lockout OFF #aa20 >
Setup Lockout ON #aa21 >
Read Setup Lockout
status
#aa22 >xx
where:
xx = 00 is OFF
xx = 01 is ON
Set Auto-On #aaB0cn1cn2x.xE-xx
where cn1 is the Ion
Gauge and cn2 is the
TC.
>
Read Auto-On #aaB1cn
where cn is Ion Gauge.
>cnx.xExx
where cn is the TC.
Table B-1 Serial Command Set (Continued)
Description Command
Response from
XGS-600 Notes