XGS-600 Gauge Controller
DRAFT 4/15/14
73
Set Setpoint ON Delay
time
#aaChx.x
where h is setpoint
number 1-8 and x.x is
delay in seconds.
C is part of the hex
command number.
Example: #00C31.2 sets
on delay to 1.2 seconds
for setpoint # 3
> Setpoint
must first be assigned to
a sensor using setpoint
screen or command 6h.
Set Setpoint Off
pressure
level
#aa7hcnx.xxxE-xx
where h is setpoint
number 1 - 8
> Setpoint ON level
must be set before OFF
level.
Set Setpoint Off Delay
time
#aaDhx.x
where h is setpoint
number 1-8 and x.x is
delay in seconds
> Setpoint must be
assigned to a sensor.
Read Setpoint On
pressure level
#aa8h
where h is setpoint
number 1 - 8
>x.xxxE-xx
Read Setpoint ON
Delay time
#aaEh
where h is setpoint
number 1-8
>x.x
where x.x is delay in
seconds
Read Setpoint Off
pressure level
#aa9h
where h is setpoint
number 1 - 8
>x.xxxE-xx
Read Setpoint Off
Delay time
#aaFh
where h is setpoint
number 1 - 8
>xx
where x.x is delay in
seconds
AUX and CNV Commands
Calibrate #aaA1cn
where cn is a CNV or
AUX
> This command has the
same function as CAL
on the main screen, i.e.
used for both
atmosphere and
vacuum calibrations.
Set Atmosphere Value #aaA3x.xxxE+xx
where pressure is local
atmospheric value.
> Same as System Setup
screen Atmosphere
Value.
Table B-1 Serial Command Set (Continued)
Description Command
Response from
XGS-600 Notes