XGS-600 Gauge Controller
DRAFT 4/15/14
79
P>MAX is reported as E09 ---- {0Eh} {00h} {09h}
Parameter information consists of two bytes, where Sensitivity is xx.xx and emission current
and gas correction are x.xxx.
For example:
Sensitivity = 20.00 --- {20h} {00h}
emission current = 3.500ma - {35h} {00h}
C.1.4 Reading All Pressure Channels
You can read all installed gauge channels with a single query command. The number of
returned bytes is three times the number of installed gauges. The gauges are read in the
order of the BCD base address. No separators are sent between bytes.
C.1.5 Read MultiGauge Contents Command
To maintain compatibility with Multigauge, which only had 5 slots, this command outputs
the ID of cards that are installed in addresses 1-5 only, even if a card is installed in BCD
address 6.
The Multigauge's Porting capability is not supported by the XGS600