XGS-600 Gauge Controller
80
DRAFT 4/
15/14
C.2 Command List
Table C-2 lists all of the Multigauge BCD commands supported by the XGS600. The
XGS600 does not support all old commands.
Table C-2 BCD Command List
Command Byte # and Syntax Byte # - Response
Read Card Contents
(1 byte)
Response
(5 bytes)
Byte 1- {01h}
where:
10h = HFIG card set for UHV24 or
UHV24p
20h = HFIG card set to 564, MBA100,
MBA200, MBA2
30h = HFIG card set to 572,571,563
3Ah= IMG card
40h = emulated 4ch TC card
48h = Convection card
FFh = empty slot
Byte 1 - {card #1-ID}
Byte 2 - {card #1-ID}
Byte 3 - {card #1-ID}
Byte 4 - {card #1-ID}
Byte 5 - {card #1-ID}
Read Pressure
(2 bytes)
Response
(3 bytes)
Byte 1- {02h)
Byte 2- {card info byte}
while
---- appears, the response is 3 bytes
of {00h}
Byte 1 - {mantissa digits 1,2}
Byte 2 - {mantissa digits 3,4}
Byte 3 - {exponent}
Read SW Revision
(1 byte)
Response
(2 bytes)
Byte 1- {05h}
where the revision is Px.y
Byte 1 - {0xh}
Byte 2- {0yh)
Reset to Default
(1 byte)
Byte 1- {06h}
Read Pressure Units
(1byte)
Response
(1 byte)
Byte 1- {13h}
where:
00h = Torr
01h = mBar
02h = Pa
Byte 1 - {xxh}
Emission OFF
(2 byte)
Byte 1- {30h}
Byte 2- {card info byte}