XGS-600 Gauge Controller
DRAFT 4/15/14
81
Emission ON
(2 bytes)
Byte 1- {31h}
Byte 2- {card info byte}
Emission ON Fil2
(2 bytes)
Byte 1- {33h}
Byte 2- {card info byte}
Read Emission Status
(2 bytes)
Response
(1 byte)
Byte 1- {32h}
Byte 2- {card info byte}
where
00h = OFF
01h = ON
02h = ON Fil2
Byte 1 - {xxh}
Degas OFF
(2 bytes)
Byte 1- {40h}
Byte 2- {card info byte}
Degas ON
(2 bytes)
Byte 1- {41h}
Byte 2- {card info byte}
Read Degas Status
(2 bytes)
Response
(1 byte)
Byte 1- {42h}
Byte 2- {card info byte}
where
00h = OFF
01h = ON
Byte 1 - {xxh}
Read Emission Curr
(2 bytes)
Response
(2 bytes)
Byte 1- {52h}
Byte 2- {card info byte}
Byte 1 - mantissa digits 1,2}
Byte 2- {mantissa digits 3,4}
Set Emission Curr
(4 bytes)
Byte 1- {53h}
Byte 2- {card info byte}
Byte 3- {mantissa digits 1,2}
Byte 4- {mantissa digits 3,4}
Read Sensitivity
(2 bytes)
Response
(2 bytes)
Byte 1- {54h}
Byte 2 - {card info byte}
Byte 1- {mantissa digits 1,2}
Byte 2- {mantissa digits 3,4}