VII. ACCESSING & EDITING THE CUSTOM SCALE VARIABLES
The Custom Scale will only be active if it is selected in Section VI.
Custom Scale Score represents the E20CP-III Commercial Scale score as modified by the equation:
(Custom Scale Score) QMS = A x (S x S) + (B x S) +/- (C)
Where:
S: Represents the E20CP-III Commercial Scale root score
A: Represents a multiplication factor applied to the squared Commercial Scale score
B: Represents a Commercial Scale score multiplier
C: Represents an add or subtract Commercial Scale score offset constant
VARIABLE DEFAULT VALUES & LIMITS
Variable A: Default (0.0000) / Limits (0.00000 to 0.59998)
Variable B: Default (1.000) / Limits (0.000 to 9.876)
Variable C: Default (00.00) / Limits (+/- 00.00 to 98.76)
Depress [FNC] to access / edit the Custom Scale variables
The display first shows: << CUSTOM – SCALE >>
QMS = A x (S x S) + (B x S) + C
Followed by the top line of the display showing:
QMS = A x (S x S) + (B x S) + C
alternating with
[ENTER] To CHANGE
alternating with
PRESS [FNC] FOR NEXT
The bottom line of the display shows: A x (S x S) A = .00000
To move to the B variable without editing the A variable displayed, depress the [FNC] key
To edit the A variable, depress the [ENTER] key
The display shows: [ENTER] TO SAVE AND EXIT
A x (S x S) A = .00000