A-MIP 200 V1.06
Installation
Identification Label
The identification label is mounted inside the Aivia.
For all communications with Customer Service or your
representative, please have the Aivia reference number
and serial number available.
Initial opening
To open the Aivia door before wall mounting, use a
small screwdriver.
Slide the screwdriver between the door and the Aivia
frame.
Lift up the screwdriver, taking care not to mark the
components.
. Do this on both sides of the Aivia.
Serial Number
Reference Options
Important
Read this manual with caution before setting up and
using the Aivia.
Never install Aivia under direct sunlight exposure.
You risk exposing the defibrillator to excessive tem-
peratures.
Aivia presentation
1 Electronic compartment.
2 Light sensor.
3 Seals passageway
4 Heating film *.
5 Night and alert indicators.
6 Door
7 Identification plate. * Option
1
2
4
3
6
5
7
9
8
8 Fixing holes.
9 AED holding sandow .