A-MIP 200 V1.06
Use
Night mode operation
The Aivia is equipped with a light detector. In low
ambient light, blinking white lights help identify the
location of the Aivia.
Take hold of the defibrillator
Go to the victim
Activate the defibrillator
Follow the defibrillator instructions
Heating option operation
The heater allows the defibrillator operating temperature
to be held within certain limits. In extreme cold, ensure
that the temperature warning lights are not blinking red.
Ventilation allows hot air within the Aivia to escape. In
hot weather, ensure that the temperature warning lights
are not blinking red.
Using the Aivia
If you are a witness of an incident requiring use of a
defibrillator.
Call emergency services.
Speak in a strong, clear voice, describe the situation and
specify the location of the incident.
Open the Aivia door by pulling the top.
The visual warning will be given by blinking red lights.
If your Aivia is equipped with an audible alarm, it will
sound for about 3 minutes.
Aivia Maintenance
Clean the surfaces of the Aivia using a soft damp cloth.
Wipe the Aivia with a soft, clean and dry cloth. Do not use
chemical products or cleaners, which may damage the
Aivia.
Do not rub the Aivia with a hard object, which may scratch
or permanently damage the surfaces.
Do not use any high pressure cleaner.
Regularly inspect seal condition
If the Aivia seals have been opened, make sure that all
items of the defibrillator are present and working properly.
In case of any doubt, contact your defibrillator distributor.
Alarm operation
Door opening alarm:
When the door is opened, blinking red lights give a
visual warning.
If your Aivia is equipped with an audible alarm
(Reference: X2A200-XX100 or X2A200-XX101) it will
sound for about 3 minutes.
Close the door to stop the alarm.
Temperature alarm:
The temperature alarm, indicated by a continuously
blinking red light when the door is closed, is
triggered below 5º or above 40º.
After using the defibrillator
• Do not place the defibrillator back inside the AIVIA be-
fore reactivating it.
• Follow the instructions in the defibrillator’s manual to
reset it for use.
• The manufacturer may be able to help reactivate the
defibrillator. Please feel free to contact them.
• After reactivating the defibrillator, replace it inside the
AIVIA using the instructions in the section "Setting up the
defibrillator."
• Replace the seal using the instructions in the section
"Installing a seal".