38 Alere Cholestech LDX
®
System User Manual
Error Message Problem Action to Take
Data file
empty
This is a normal
message
that appears:
1. When the optics check
has been run and DATA
button is pushed.
2. A cassette has been run,
the power supply has been
unplugged from the analyzer,
and DATA button is pushed.
3. When an error message
occurs, i.e., “Used
Cassette,” and DATA
button is pushed.
End of data
info
This is a normal
message that
appears:
1. CHD risk is off and
DATA button is pushed
twice after results appear.
2. Framingham risk is on and
the STOP button is pushed
when “Risk? – RUN=yes,
STOP=no” appears.
Expired
Cassette
Cassette
expiration date,
month has been
reached.
Check calendar in the
analyzer to make sure it
is set correctly. Check
expiration date on cassette
pouch or box.
Low Battery
Internal battery
voltage is too low.
Call Alere. Internal battery
is not user serviceable.
Mag Read
Error
See User
Manual
Unable to read
magnetic stripe on
the cassette.
1. Wipe magnetic stripe
with a soft tissue and
restart test. If second test
fails, repeat test using a new
cassette.
2. Check cassette position;
magnetic stripe must be on
the right.
3. If message reappears,
call Alere.
RUN button
pushed without
cassette in drawer
Place cassette in drawer
before pushing RUN
button.

Other manuals for Alere Cholestech LDX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Alere Cholestech LDX and is the answer not in the manual?

Alere Cholestech LDX Specifications

General IconGeneral
BrandAlere
ModelCholestech LDX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals