39Alere Cholestech LDX
®
System User Manual
Error Message Problem Action to Take
Optics Test
Fail
See User
Manual
A problem with
the optical system
of the analyzer
or a faulty Optics
Check Cassette.
Repeat optics check with
another Optics Check
Cassette. If the message
reappears, call Alere.
The analyzer will now be
deactivated.
Reaction Did
Not Occur
(nnnn)
Insufcient
sample volume
was applied to
the cassette due
to air bubbles or
incompletely lled
capillary tube or
pipetting error.
1. Repeat test with a
new cassette and new
ngerstick sample.
2. Repeat test with a new
pipette tip and cassette.
Be sure pipette tip is rmly
in place.
Insufcient sample
volume reached
reaction pads due
to high hematocrit
of sample.
Repeat test with a new
cassette and new ngerstick
sample. If message
reappears, test patient by
alternate method.
Improper sample
collection: Sample
clotted because it
was left in capillary
tube too long.
Repeat test with a
new cassette and new
ngerstick sample. Pay
close attention to proper
technique.
Improper sample
collection: Finger
was milked to
obtain sample, or
capillary tube was
lled too slowly.
Repeat test with a
new cassette and new
ngerstick sample. Pay
close attention to proper
technique.
The cassette is
malfunctioning.
Repeat test with a new
cassette.
RH Sensor
Fail
The humidity
sensor has
become
disconnected or
has failed.
Call Alere.