The device checks that the generator
frequency and voltage lie
the mentioned limits, the device carries out all the adjusting
operations necessary to make them lie within the
defined. The 'parallel breaker closing' command will
3.6 Main and spare settings
The device can store two groups of settings, namely
spare', which can be selected through keyboard, PC or
The group of setting being used is shown on the display as
to use the SPARE settings. This 'forcing' can be carried out from
the network or by permanently enabling the digital input (IN
characters <!> which appear near the group already selected;
when the input is disabled, the group of settings previously
selected is restored and the characters <!> disappear (see
Diagnostic activities are envisaged which are automatically carried
out upon powering on the device as well as from time to time
during the running, and which allow the following actions to be
, turns off the green LED (running device)
and turns on the red LED ( faulty devi
timed operation (VITAL FAULTS)
The following faults are defined as NOT
The following faults are defined as VITAL FAULTS:
vital faults the device automatically displays the
code relative to the kind of fault occ
urred; if allowed by the device,
this condition can only be solved by enabling the 'settings change'
In the event that, upon starting up the equipment, an EEPROM
failure makes it impossible to retrieve the information on the
selected dialogue language, the device will continue on
using Italian as default language.
The device includes the 'Watch Dog'.
The diagnostic activity is also aimed at signalling any condition of
'inhibited equipment': the relevant code is display
relay is de-energized and the red LED (faulty device) lights on.
The diagnostic menu also indicates the date when last
The output circuits are composed of six relays: each of them is
provided with two contacts (see
Each remaining relay, if duly programmed from the keyboard, PC
e six relays is of monostable type:
it allows to immediate return to
the quiescent position when the
cause that operated it is
The output relays, except for the diagnostic relays, can all be
to 'normally energized' (NORMA
or to 'normally de-energized' (NORMAL.ON=OFF): the selection
occurs bymeans of the S1 jumper which is
relay has been pulled out (see fig. 2a,