D-000348'Rev.'A'
49'
Appendix A: Via400 Specifications
MODEL Via400, 400m, 400x
• Fits individuals from 4’8” (142 cm) to 6’4” (193 cm); 18.5” (47 cm) hip width, 58” (147 cm)
hip circumference.
PERFORMANCE
• User weight capacity: 85 - 400 pounds (181.4 kg)
• Body Weight Range Adjustment:
o Up to 320 lbs (145.2 kg) 20% – 100% of user’s body weight
o >320 lbs (145.2 kg) 35% - 100% of user’s body weight
• Running surface area:
- 20 inches (51 cm) wide
- 58 inches (147 cm) long
• Speed range:
o 400/400m
• Forward: 0 – 12 mph (19.3 km/hr)
• Reverse: 3 mph (4.8 km/hr)
o 400x
§ Forward: 0 – 15 mph (24.1 km/hr)
§ Reverse: 5 mph (8.0 km/hr)
• Elevation: 0 - 15%
DIMENSIONS
• Length: 84 inches (213 cm)
• Width: 38 inches (96.5 cm)
• Height: 74 inches (188 cm)
• Weight: 750 pounds (340 kg), approximately
• Step-up height: 8” (20 cm)
RECOMMENDED ROOM DIMENSIONS
• Provide a footprint at least 12ft (3.66 meters) long by 8ft (2.44 meters) wide for
adequate spacing around the machine
• A minimum 8 foot (2.44 meters) ceiling height is recommended
ELECTRICAL
• Power Requirements: 220 VAC 20A, 50/60 Hertz, NEMA 6-20R receptacle
• Locate the front of the system within 10ft (3.05 meters) of the electrical outlet.
ENVIRONMENTAL
Operating Conditions:
• Ambient Temperature: 55°F to +84°F (+13°C to +29°C)
• Relative Humidity: 20 to 95%
Transportation & Storage Conditions:
• Temperature Range: 0°F to +120°F (0°C to +49°C)
• Relative Humidity: 20 to 95%f'
• '

Table of Contents