ENGLISH 9
Guidance and manufacturer’s declaration – electromagnetic immunity
IMMUNITY test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601 test
level
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
Compliance level
3 Vrms
3 V/m
Electromagnetic environment – guidance
Portable and mobile RF communications
EQUIPMENTSHOULD
be used no closer to any part of the Ameda Elite Breast
Pump, including cables, than the recommended
SEPARATIONDISTANCECALCULATEDFROMTHEEQUATION
APPLICABLETOTHEFREQUENCYOFTHETRANSMITTER
Recommended separation distance
d = 1,2
P
d = 1,2
P
80 MHz to 800 MHz
d = 2,3
P
800 MHz to 2,5 GHz
Where
P
is the maximum output power rating of the
TRANSMITTERINWATTS7ACCORDINGTOTHETRANSMITTER
manufacturer and d is the recommended separation
DISTANCEINMETERSM
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey,
a
should be less than
THECOMPLIANCELEVELINEACHFREQUENCYRANGE
b
)NTERFERENCEMAYOCCURINTHEVICINITYOFEQUIPMENT
marked with the following symbol:
The Ameda Elite Breast Pump is intended for use in
the electromagnetic environment specified below.
The customer or the user of the Ameda Elite Breast Pump
should assure that it is used in such an environment.
./4% !T-(ZAND-(ZTHEHIGHERFREQUENCYRANGE
applies.
./4% 4HESEGUIDELINESMAYNOTAPPLYINALLSITUATIONS
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for
RADIOCELLULARCORDLESSTELEPHONESANDLANDMOBILERADIOSAMATEUR
radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnet-
ic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the
location in which the Ameda Elite Breast Pump is used exceeds the
applicable RF compliance level above, the Ameda Elite Breast Pump
should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary,
such as re-orienting or relocating the Ameda Elite Breast Pump.
b
/VERTHEFREQUENCYRANGEK(ZTO-(ZlELDSTRENGTHS
should be less than 3 V/m.
Recommended separation distances between portable and mobile RF
communications equipment and the Ameda Elite Breast Pump
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
150 kHz to 80 MHz
d = 1,2
P
0,12
0,38
1,2
3,8
12
80 MHz to 800 MHz
d = 1,2
P
0,12
0,38
1,2
3,8
12
800 MHz to 2,5 GHz
d = 2,3
P
0,23
0,73
2,3
7,3
23
Separation distance according to frequency of transmitter
m
The Ameda Elite Breast Pump is intended for use
in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of
the Ameda Elite Breast Pump can help prevent
electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications
EQUIPMENTTRANSMITTERSANDTHE!MEDA%LITE"REAST0UMP
as recommended below, according to the maximum output
POWEROFTHECOMMUNICATIONSEQUIPMENT
For transmitters rated at a maximum output power not
listed above, the recommended separation distance d in
METERSMCANBEESTIMATEDUSINGTHEEQUATIONAPPLICABLE
TOTHEFREQUENCYOFTHETRANSMITTERWHERE
P
is the
maximum output power rating of the transmitter in
WATTS7ACCORDINGTOTHETRANSMITTERMANUFACTURER
./4% !T-(ZAND-(ZTHESEPARATIONDISTANCEFORTHE
HIGHERFREQUENCYRANGEAPPLIES
./4% 4HESEGUIDELINESMAYNOTAPPLYINALLSITUATIONS
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.

Other manuals for Ameda Elite

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ameda Elite and is the answer not in the manual?

Ameda Elite Specifications

General IconGeneral
Suction LevelsAdjustable
Cycle SpeedAdjustable
BPA FreeYes
Double or Single PumpingDouble
Battery OperatedNo
Closed SystemYes
Warranty1 year
Power SourceAC Adapter

Related product manuals