Guidance and manufacturer’s declaration – electromagnetic immunity
IMMUNITY test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601 test
level
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2,5 GHz
Compliance level
3 Vrms
3 V/m
Electromagnetic environment – guidance
Portable and mobile RF communications
EQUIPMENTSHOULD
be used no closer to any part of the Ameda Elite Breast
Pump, including cables, than the recommended
SEPARATIONDISTANCECALCULATEDFROMTHEEQUATION
APPLICABLETOTHEFREQUENCYOFTHETRANSMITTER
Recommended separation distance
d = 1,2√
P
d = 1,2√
P
80 MHz to 800 MHz
d = 2,3√
P
800 MHz to 2,5 GHz
Where
P
is the maximum output power rating of the
TRANSMITTERINWATTS7ACCORDINGTOTHETRANSMITTER
manufacturer and d is the recommended separation
DISTANCEINMETERSM
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey,
a
should be less than
THECOMPLIANCELEVELINEACHFREQUENCYRANGE
b
)NTERFERENCEMAYOCCURINTHEVICINITYOFEQUIPMENT
marked with the following symbol:
The Ameda Elite Breast Pump is intended for use in
the electromagnetic environment specified below.
The customer or the user of the Ameda Elite Breast Pump
should assure that it is used in such an environment.
./4% !T-(ZAND-(ZTHEHIGHERFREQUENCYRANGE
applies.
./4% 4HESEGUIDELINESMAYNOTAPPLYINALLSITUATIONS
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for
RADIOCELLULARCORDLESSTELEPHONESANDLANDMOBILERADIOSAMATEUR
radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnet-
ic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the
location in which the Ameda Elite Breast Pump is used exceeds the
applicable RF compliance level above, the Ameda Elite Breast Pump
should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary,
such as re-orienting or relocating the Ameda Elite Breast Pump.
b
/VERTHEFREQUENCYRANGEK(ZTO-(ZlELDSTRENGTHS
should be less than 3 V/m.
Recommended separation distances between portable and mobile RF
communications equipment and the Ameda Elite Breast Pump
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
150 kHz to 80 MHz
d = 1,2√
P
0,12
0,38
1,2
3,8
12
80 MHz to 800 MHz
d = 1,2√
P
0,12
0,38
1,2
3,8
12
800 MHz to 2,5 GHz
d = 2,3√
P
0,23
0,73
2,3
7,3
23
Separation distance according to frequency of transmitter
m
The Ameda Elite Breast Pump is intended for use
in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of
the Ameda Elite Breast Pump can help prevent
electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications
EQUIPMENTTRANSMITTERSANDTHE!MEDA%LITE"REAST0UMP
as recommended below, according to the maximum output
POWEROFTHECOMMUNICATIONSEQUIPMENT
For transmitters rated at a maximum output power not
listed above, the recommended separation distance d in
METERSMCANBEESTIMATEDUSINGTHEEQUATIONAPPLICABLE
TOTHEFREQUENCYOFTHETRANSMITTERWHERE
P
is the
maximum output power rating of the transmitter in
WATTS7ACCORDINGTOTHETRANSMITTERMANUFACTURER
./4% !T-(ZAND-(ZTHESEPARATIONDISTANCEFORTHE
HIGHERFREQUENCYRANGEAPPLIES
./4% 4HESEGUIDELINESMAYNOTAPPLYINALLSITUATIONS
Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.