28   18-09-2007  122770 06 
 
Error:  Switch test does not work. 
Error message: 
- 
 
Likely cause:  Solution: 
Defective Main PCB  Check signal from Switch Test Input. The 
voltage between the terminals on the panel 
should be at least 4.0V.  
Check the signals SW-TEST(+) and SW-
TEST(-) on the connector CN1 on the 
INPUT PCB. 
Otherwise, replace the Main PCB and 
adjust and re-calibrate the unit 
 
Error:  Fan does not work. 
Error message: 
- 
 
Likely cause:  Solution: 
Fan obstructions  Remove any obstructions.