EasyManuals Logo

Ametek JOFRA RTC-156 A User Manual

Ametek JOFRA RTC-156 A
173 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #163 background imageLoading...
Page #163 background image
162 2014-07-30 127139 13
INPUT SPECIFICATIONS
ACCURACY IN C/F
Accuracy RTD
Pt50(90)391: -200°C to 1100°C
* Pt50(68)391: -200°C to 1100°C
(excluding sensor accuracy)
±0.038°C(±0.067°F) @ -200°C(-328°F)
±0.042°C(±0.076°F) @ -90°C(-130°F)
±0.044°C(±0.079°F) @ -50°C(-58°F)
±0.046°C(±0.082°F) @ 0°C(32°F)
±0.051°C(±0.092°F) @ 155°C(311°F)
±0.057°C(±0.102°F) @ 320°C(608°F)
±0.061°C(±0.109°F) @ 420°C(788°F)
±0.071°C(±0.128°F) @ 660°C(1220°F)
±0.073°C(±0.132°F) @ 700°C(1292°F)
±0.081°C(±0.145°F) @ 850°C(1562°F)
±0.095°C(±0.170°F) @ 1100°C(2012°F)
Accuracy RTD
Pt100(90)392: -200°C to 630°C
(excluding sensor accuracy)
±0.020°C(±0.036°F) @ -200°C(-328°F)
±0.023°C(±0.042°F) @ -90°C(-130°F)
±0.024°C(±0.042°F) @ -50°C(-58°F)
±0.024°C(±0.046°F) @ 0°C(32°F)
±0.030°C(±0.054°F) @ 155°C(311°F)
±0.035°C(±0.062°F) @ 320°C(608°F)
±0.038°C(±0.068°F) @ 420°C(788°F)
±0.045°C(±0.081°F) @ 630°C(1166°F)
Accuracy RTD
M100(90)428: -200°C to 200°C
* M100(68)428: -200°C to 200°C
(excluding sensor accuracy)
±0.020°C(±0.035°F) @ -200°C(-328°F)
±0.021°C(±0.038°F) @ -90°C(-130°F)
±0.022°C(±0.040°F) @ -50°C(-58°F)
±0.024°C(±0.043°F) @ 0°C(32°F)
±0.027°C(±0.048°F) @ 155°C(311°F)
±0.028°C(±0.050°F) @ 200°C(392°F)

Table of Contents