EasyManuals Logo

Ametek JOFRA RTC-156 A User Manual

Ametek JOFRA RTC-156 A
173 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #164 background imageLoading...
Page #164 background image
127139 13 2014-07-30 163
INPUT SPECIFICATIONS
ACCURACY IN C/F
Accuracy RTD
M50(90)428 : -200°C to 200°C
* M50(68)428 : -200°C to 200°C
(excluding sensor accuracy)
±0.038°C(±0.069°F) @ -200°C(-328°F)
±0.040°C(±0.071°F) @ -90°C(-130°F)
±0.041°C(±0.073°F) @ -50°C(-58°F)
±0.043°C(±0.076°F) @ 0°C(32°F)
±0.046°C(±0.082°F) @ 155°C(311°F)
±0.047°C(±0.083°F) @ 200°C(392°F)
Accuracy RTD
* M100(90)426: -5C to 200°C
(excluding sensor accuracy)
±0.023°C(±0.041°F) @ -50°C(-58°F)
±0.024°C(±0.043°F) @ 0°C(32°F)
±0.027°C(±0.048°F) @ 155°C(311°F)
±0.028°C(±0.050°F) @ 200°C(392°F)
Accuracy RTD
* M53(68)426: -5C to 200°C
(excluding sensor accuracy)
±0.040°C(±0.071°F) @ -50°C(-58°F)
±0.041°C(±0.073°F) @ 0°C(32°F)
±0.044°C(±0.078°F) @ 155°C(311°F)
±0.045°C(±0.080°F) @ 200°C(392°F)
Accuracy RTD
* M50(90)426: -5C to 200°C
(excluding sensor accuracy)
±0.042°C(±0.075°F) @ -50°C(-58°F)
±0.043°C(±0.077°F) @ 0°C(32°F)
±0.046°C(±0.082°F) @ 155°C(311°F)
±0.047°C(±0.084°F) @ 200°C(392°F)
Accuracy RTD
* H100(90)617: -6C to 180°C
(excluding sensor accuracy)
±0.021°C(±0.037°F) @ -60°C(-76°F)
±0.019°C(±0.033°F) @ 0°C(32°F)
±0.017°C(±0.030°F) @ 100°C(212°F)
±0.015°C(±0.027°F) @ 180°C(392°F)
Accuracy RTD
H120(90)672: -8C to 260°C
(excluding sensor accuracy)
±0.016°C(±0.029°F) @ -80°C(--112°F)
±0.015°C(±0.027°F) @ 0°C(32°F)
±0.014°C(±0.025°F) @ 100°C(212°F)
±0.012°C(±0.021°F) @ 260°C(500°F)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek JOFRA RTC-156 A and is the answer not in the manual?

Ametek JOFRA RTC-156 A Specifications

General IconGeneral
BrandAmetek
ModelJOFRA RTC-156 A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals