No Output Alarm Indication
Resolution(s):
Verify that the alarm and alarm delay setpoints are correct
Confirm the Alarm Delay or Alarm Hold Off setting is correct
Check that the output wires are properly stripped and connected at their correct positions at
their respective terminals
Verify that the alarms on the Analyzer are properly configured using the COMMAND
CENTER Software (see pages 32 – 35)
'Err' Flashes on the LCD
Resolution(s):
Look up the Error Code on page 44 and troubleshoot/resolve it
If you cannot resolve, contact AMI for further support
Analyzer Refuses to Accept Front Panel Settings
Resolution(s):
Use the COMMAND CENTER Software to verify that the Security Settings match your
preference
Analyzer Reads Zero
Resolution(s):
Hook up a tank of Span Gas and confirm the Analyzer responds upscale accurately.
No Voltage or Current Output to Recording Device
Resolution(s):
Check that the output wires are properly stripped and connected at their correct positions at
their respective terminals
Display Pressure Reading Not Correct
Resolution(s):
Perform the Initiation of the Pressure Sensor Procedure on page 15