No Output Alarm Indication
Resolution(s):
• Verify that the alarm and alarm delay setpoints are correct
• Confirm the Alarm Delay or Alarm Hold Off setting is correct
• Check that the output wires are properly stripped and connected at their correct positions at
their respective terminals
• Verify that the alarms on the Analyzer are properly configured using the COMMAND
CENTER Software (see pages 32 – 35)
'Err' Flashes on the LCD
Resolution(s):
• Look up the Error Code on page 44 and troubleshoot/resolve it
• If you cannot resolve, contact AMI for further support
Analyzer Refuses to Accept Front Panel Settings
Resolution(s):
• Use the COMMAND CENTER Software to verify that the Security Settings match your
preference
Analyzer Reads Zero
Resolution(s):
• Hook up a tank of Span Gas and confirm the Analyzer responds upscale accurately.
No Voltage or Current Output to Recording Device
Resolution(s):
• Check that the output wires are properly stripped and connected at their correct positions at
their respective terminals
Display Pressure Reading Not Correct
Resolution(s):
• Perform the Initiation of the Pressure Sensor Procedure on page 15
45