To replace the fuse:
1. Remove the 2 rear-case screws.
2. Separate the case.
3. Remove and replace the 2 A fuse (15XP-B or 35XP-A)
or 0.315 fuse (5XP-A).
4. Reassemble the meter.
Fuse:
Fast Blow 2 A/1000 V, minimum interrupt rating 30 kA
(6 x 32 mm) (Amprobe® FP200).
Fast Blow 0.315 A/1000 V minimum interrupt rating 30 kA
(6.3 x 32 mm) (Amprobe® FP300)
SPECIFICATIONS
Display:
5XP-A and 15XP-B:3 ½ digit liquid crystal display (LCD) with a
maximum reading of 1999.
35XP-A:
3 ¾ digit liquid crystal display (LCD) with a maximum
reading of 3999.
Polarity: Automatic, positive implied, negative polarity
indication.
Overrange:
0o or -0o is displayed.
Battery life:
Approximately 200 hours typical with carbon-
zinc battery. 400 hours with alkaline battery.
Low battery indication: The
symbol is displayed when
the battery voltage drops below the operating level.
Operating environment: 0 °C to 50 °C at <70 % R.H.
Power: Single standard 9 V battery, NEDA 1604, JIS 006P, IEC
6F22.
Dimensions: 155 mm (H) ×72 mm (W) ×32 mm (D)
Weight: Approximately 210 g including battery.