EasyManua.ls Logo

Amprobe THWD-5 - Calibration; Troubleshooting; Power On But No Display; Flickering Display and Then Disappears

Amprobe THWD-5
8 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6
CALIBRATION
Make sure the THWD-5 is turned off. Plug the meter’s probe into a known trusted 32.8% 1.
salt bottle. Press ”On/Off + F/C” pushbuttons simultaneously for more than one second to
enter the calibration mode.
3x.x% will be flashing on LCD (see fig.1). After 30 minutes the ”3x.x%” will stop flashing. 2.
The calibration for the ”3x.x%” is now completed. Proceed with the 7x.x% calibration.
Fig. 1
Remove the meter from the 32.8% salt bottle and place it into a 75.3% salt bottle.3.
Press Min Max Reset pushbutton for more than one second to enter the7x.x% calibration
mode. ”7x.x%” will be flashing on LCD. After 30 minutes the ”7x.x%” will stop flashing.
The calibration process is now complete.
Note:
Auto power off is automatically disabled during calibration procedure.•
In order to get high accuracy, you must do the calibration at a 23 • ± 2°C (73 ± 2°F) and steady
airflow in the room.
If the value differs from 7x.x% or 3x.x% for more than 0.3%, the calibration fails .(see •
trouble-shooting)
TROUBLESHOOTING
Power on but no display
Make sure the time of pressing the On/Off pushbutton is more than 100 ms.•
Check the batteries are in place, good contact and correct polarity.•
Replace the batteries and try again.•
Flickering display and then disappears
Check whether the low battery indicator displayed before display disappeared. If yes, •
replace the batteries.
Turn on the meter by pressing ”• On/Off + Hold” key to disable auto power off.

Related product manuals