GPIB Operation HP438A Emulation Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-125
FH (average hold)
FM (Set average value)
GT (Set group trigger)
Syntax: FH
Remarks: Hold filter sets the filter mode to Manual from Auto, but retains the
auto filter setting. This function is the same as the AVGM command.
Related
commands:
FM, FH, FA
Syntax: FM<val>EN
val: 0 to 9
Remarks: Sets the filter length for the averaging of sensor data. For HP438A
emulation, the filter length is defined as the number 2 to the power of
<val>. For example, the command FM5EN would be 2
5
, or 32.
Related
commands:
FH, FA
Syntax: GT<mode>
mode:
0
1
2
Remarks: 0 = Ignore Group Execute Trigger (GET) command
1 = Trigger immediate response to ‘GET’ command
2 = Trigger with delayed response to ‘GET’ command
The GTn command configures what the device does when it receives the
‘GET’ command.
For example: GT1 sets the’ GET’ (Group Execute Trigger) to perform a
TR1 type trigger.
Related
commands:
TR