EasyManua.ls Logo

Anritsu ML2438A - FH (Average Hold)

Anritsu ML2438A
314 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GPIB Operation HP438A Emulation Commands
ML2437A/38A OM/PM PN: 10585-00001 Rev. P 8-125
FH (average hold)
FM (Set average value)
GT (Set group trigger)
Syntax: FH
Remarks: Hold filter sets the filter mode to Manual from Auto, but retains the
auto filter setting. This function is the same as the AVGM command.
Related
commands:
FM, FH, FA
Syntax: FM<val>EN
val: 0 to 9
Remarks: Sets the filter length for the averaging of sensor data. For HP438A
emulation, the filter length is defined as the number 2 to the power of
<val>. For example, the command FM5EN would be 2
5
, or 32.
Related
commands:
FH, FA
Syntax: GT<mode>
mode:
0
1
2
Remarks: 0 = Ignore Group Execute Trigger (GET) command
1 = Trigger immediate response to ‘GET’ command
2 = Trigger with delayed response to ‘GET’ command
The GTn command configures what the device does when it receives the
‘GET’ command.
For example: GT1 sets the’ GET’ (Group Execute Trigger) to perform a
TR1 type trigger.
Related
commands:
TR

Table of Contents