2.21 Setting External Mixer
2-997
2
SCPI Device Message Details
[:SENSe]:MIXer:LOSS[:FIXed]?
External Mixer Conversion Loss Fixed Value Query
Function
This command queries the external mixer's fixed conversion loss.
Query
[:SENSe]:MIXer:LOSS[:FIXed]?
Response
<power>
Parameter
<power>
Conversion Loss of External Mixer
Range 0.00 to 99.99 dB
Resolution 0.01 dB
Suffix code None. Value is returned in dB units.
Details
This function is available when Option 044/045 is installed for MS2830A.
This function is available when Option 044/046 is installed for MS2840A.
One value is held per external mixer band.
Example of Use
To query the current external mixer's fixed conversion loss.
MIX:LOSS?
> 10.00