Chapter
Native Device Message Details
4-61
4
Native Device Message Details
CALC:SPUR:LIM:ABS:DATA:STOP/CALC:SPUR:LIM:ABS:DATA:STOP?
Spurious Emission Limit Stop Level
Function
This command sets the absolute level limit of the stop frequency of each
segment for the Spurious Emission measurement.
Command
CALC:SPUR:LIM:ABS:DATA:STOP
ampl_1,ampl_2,ampl_3,ampl_4,ampl_5,ampl_6,ampl_7,ampl_8,
ampl_9,ampl_10,ampl_11,ampl_12,ampl_13,ampl_14,ampl_15,a
mpl_16,ampl_17,ampl_18,ampl_19,ampl_20
Query
CALC:SPUR:LIM:ABS:DATA:STOP?
Response
ampl_1,ampl_2,ampl_3,ampl_4,ampl_5,ampl_6,ampl_7,ampl_8,
ampl_9,ampl_10,ampl_11,ampl_12,ampl_13,ampl_14,ampl_15,a
mpl_16,ampl_17,ampl_18,ampl_19,ampl_20
Value is returned in dBm units, without suffix code.
Parameter
ampl_n
Absolute level limit of stop frequency in Segment
n
Range 200 dBm to 50 dBm
Resolution 0.01 dB
Suffix code DBM,DM
Example of Use
To set the absolute level limit of the stop frequency in each segment.
CALC:SPUR:LIM:ABS:DATA:STOP
-13,-13,-13,-13,-13,-13,-10,-10,-13,-13,-13,-13,-13,-13,
-10,-10,-13,-13,-13,-13
CALC:SPUR:LIM:ABS:DATA:STOP?
>
-13.00,-13.00,-13.00,-13.00,-13.00,-13.00,-10.00,-10.00,
-13.00,-13.00,-13.00,-13.00,-13.00,-13.00,-10.00,-10.00,
-13.00,-13.00,-13.00,-13.00