Chapter 2 SCPI Device Message Details
2-52
[:SENSe]:FREQuency:DOMain:COUPle ON|OFF|1|0
Couple Time/Freq. Domain
Function
Sets whether to couple the time domain parameters and frequency
domain parameters.
Command
[:SENSe]:FREQuency:DOMain:COUPle <switch>
Parameter
<switch>
Enables/disables coupling
ON|1
Couples (shares) the RBW, VBW, detection mode,
and trace point between the time domain and
the frequency domain (Default).
OFF|0
Separates the RBW, VBW, detection mode, and
trace point for the time domain from those for
the frequency domain.
Example of Use
To disable coupling.
FREQ:DOM:COUP OFF
[:SENSe]:FREQuency:DOMain:COUPle?
Couple Time/Freq. Domain Query
Function
Queries whether the time domain parameters and frequency domain
parameters are coupled.
Query
[:SENSe]:FREQuency:DOMain:COUPle?
Response
<switch>