Pure2 User Manual 23
10M/MX Calibration
The Pure2 can be calibrated periodically in order to calibrate the accuracy of the internal
clock by connecting it to the Antelope Audio 10M/MX Atomic Clock. This is not something
that necessarily needs to be done very often, but if desired, please follow these
instructions:
Connect a 10M/MX unit to Pure2 via the dedicated 10M input.
Power on the 10M and wait until it reaches operating temperature.
With the Pure2 powered up, press and hold the Power button.
Turn the Rotary control to navigate to the ‘10M Calibration’ option.
Confirm the selection by pressing the Rotary control.
The Calibration screen appears showing the Oven Temperature and AC Lock
status.
Wait for the sign Locked. Error value will now be displayed at the bottom of the
screen.
Wait for the calibration to take place.
When the Err. Number becomes stable, press the Rotary control (7) again to end
the calibration.
You will see “DONE” displayed on the LCD screen. You may exit the menu pressing
the Volume Select button.