2.3 AUTO OFF SETUP
When scale is on and at zero and stable,
long press UNIT key enter the auto
off setup, use ZERO key to enable
or disable, then press PRINT key to
confirm and back to weighing mode.
on enable (scale will turn off after 10 mins haven’t been use; to be able to use the scale, needs to turn
on the on/ff switch)
off disable
2.4 DIVISION SETUP
When scale is on and at zero and stable, long press FUNCTION key enter the division setup, use
ZERO key to select the divisions, then press PRINT key to confirm and back to weighing mode.
-301_Calibration
A. Setting up division, decimal point, max. capacity and calibration
Press and hold “PRINT” key,then turn on the scale,to enter division setting:
1. Division Setting:
LCD shows [d x] then press “ZERO” key to select 1、2、5、10、20、50,and then
press “PRINT” key to confirm,and enter the decimal point setting.
Ex. When display shows [d 5],Press “PRINT” key,the division set as 5,then enter the
decimal point setting.
2. Decimal point setting:
LCD shows [p x] then press “ZERO” key to select 0、0.0、0.00、0.000、
0.0000,P.00000,and then press “PRINT” key to confirm,and enter the maximum capacity
setting.
8