2.3 AUTO OFF SETUP
When scale is on and at zero and stable,
long press UNIT key enter the auto
off setup, use ZERO key to enable
or disable, then press PRINT key to
confirm and back to weighing mode.
on enable (scale will turn off after 10 mins haven’t been use; to be able to use the scale, needs to turn
on the on/ff switch)
off disable
2.4 DIVISION SETUP
When scale is on and at zero and stable, long press FUNCTION key enter the division setup, use
ZERO key to select the divisions, then press PRINT key to confirm and back to weighing mode.
-301_Calibration
A. Setting up division, decimal point, max. capacity and calibration
Press and hold PRINT” keythen turn on the scaleto enter division setting:
1. Division Setting
LCD shows [d x] then press ZEROkey to select 125102050and then
press “PRINTkey to confirmand enter the decimal point setting.
Ex. When display shows [d 5]Press “PRINT” keythe division set as 5then enter the
decimal point setting.
2. Decimal point setting
LCD shows [p x] then press ZEROkey to select 00.00.000.000
0.0000P.00000and then press PRINTkey to confirmand enter the maximum capacity
setting.
8