- 4 -
5 Conductivity Measurement
1) Short press to turn on tester. Rinse probe in distilled water and dry it.
2) Stir probe in the sample solution gently, leave it to stand. Get readings after the smile icon
comes up and stays.
3) Press to switch from Conductivity to TDS or salinity.
6 Notes
1) The tester adopts 84 µS (sold separately), 1413 µS and 12.88 mS standard calibration
solutions. Users can use 1-3 point calibrations as needed. For most circumstances,
calibrating in 1413 µS to complete 1st point calibration will meet testing requirements.
Calibration Icons
Measuring Range
0 to 200 µS/cm
200 to 2000 µS/cm
2 to 20 mS/cm
2) The tester has already been calibrated after manufacture. Usually users can use the
tester right away, or test it in the calibration solutions to test its accuracy. When error is
large, calibrate it before using.
3) The factory default temperature compensation coefficient is 2.0%/˚C. Users can adjust it
in P3 parameter setting according to the following table or laboratory data.
Solution
Temp. Compensation Coefficient
NaCl
2.12%/˚C
5% NaOH
1.72%/˚C
Diluted Ammonia
1.88%/˚C
10% Hydrochloric Acid
1.32%/˚C
5% Sulfuric Acid
0.96%/˚C

Other manuals for Apera Instruments EC60

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Apera Instruments EC60 and is the answer not in the manual?

Apera Instruments EC60 Specifications

General IconGeneral
BrandApera Instruments
ModelEC60
CategoryTest Equipment
LanguageEnglish

Related product manuals