iCH Auto/English 19
1. Water chamber is not properly
assembled.
2. Water chamber worn out.
1. Remove the water chamber from
the heated humidifier, pour out
the water and reassemble the
water chamber again, making
sure water chamber closes
securely, and fill the water till it
reaches fill line and check if it still
leaks or not.
2. Replace with a new water
chamber.
Error / Warning Messages shown on LCD
Error:
Primary function can’t
execute.
Error for abnormal system settings
Error for flow generator failure
Error for abnormal timer setting or timer
failure
Error for flow sensor failure
Error for heater failure (thermal fuse
blows off)
Error for pressure sensor failure
System memory is nearly full
SD card module communication failure
Remove the SD card while data is being
processed
Inconsistent ID no. of device with SD
card
NOTE: When the warning message appears, contact your doctor or equipment provider to
download the memory data and reset the meter.