EasyManua.ls Logo

Apex Digital XT Fit - Technical Specifications; Device Specifications; Environmental and Classification Details

Apex Digital XT Fit
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14
English
English
8. Technical Specifications
Item
Specifications
Mode of Operation
Continuous
Pressure Range
4 20 cmH
2
O (adjustable in 0.5 cmH
2
O increment)
Ramp Time
0 45 minutes (adjustable in 5-minute increment)
Ramp Starting Pressure
3 to therapy pressure
Altitude Compensation
(Manual setting)
Level 1~8 for 0 ~ 8000 ft when the pressure set at 4-18 cmH
2
O.
Level 1~5 for 0 ~ 5000 ft when the pressure set at 18.5-20 cmH
2
O
Dimensions (W x D x H)
14.5 x 13.0 x 10.0 cm or 5.7” x 5.1” x 3.9”
Weight
Approximately 800 g or 1.76 lb
Sound Level
28 dBA at 10 cmH
2
O, 1 meter distance
Power Input
AC100-240V, 50/60 Hz, 0.5-0.3A
Environment
Temperature
Operating: +5° C to +35° C (+41° F to +95° F)
Storage/Transport: -15º C to +50º C (+5° F to +122° F)
Humidity
Operating: 15%RH to 95%RH non-condensing
Storage/Transport: 10%RH to 90%RH non-condensing
Air Tubing
Flexible plastic, 1.8m (approx.)
Classification:
Class II
Type BF, Applied Parts Nasal Mask.
Not suitable for use in the presence of a flammable anesthetic
mixture.
IP21: Protected against solid foreign objects of 12.5 mm and
greater; Protected against vertically falling water drops.
Continuous operation.Continuous operation.
Note: The manufacturer reserves the right to modify the specification without notice.

Other manuals for Apex Digital XT Fit

Related product manuals