EasyManua.ls Logo

AR PL7004 - Page 23

AR PL7004
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The FI7000 interface provides both power to
and communication with the FL/PL 7000 series
probes. A low-level loop back fiber optic con-
nection is used to sense unexpected disconnect
of laser driven fibers to ensure user-safe opera-
tion. As an additional safety measure, shuttered
connectors are used on all fiber optic connec-
tions to and from the probe.
The FI7000 can be used with the FM7004A Field
Monitor (using the provided 0.5 meter fiber optic
cable) or a PC by direct communication using
the USB, GPIB, or RS-232 interfaces.
Features
Specifications
FI7000
Probe Interface
AR RF/Microwave
Instrumentation
160 School House Rd
Souderton, PA 18964
215-723-8181
For an applications engi-
neer call:800.933.8181
www.arworld.us
To order AR Products, call 215.723.8181. For an applications engineer call:800.933.8181. Direct to Service call: 215.723.0275 or
email: service@arworld.us For Faxing Orders:866.859.0582 (Orders Only Please) info@arworld.us
Approved for public release by AR RF/Microwave Instrumentation 020816
Interfaces:
IEEE-488 (GPIB)
USB 2.0, Test and Measurement class
RS-232 (19200 Baud)
Fiber Optic Serial (19200 Baud) reserved for con-
nection to FM7004A Field Monitor
Laser Fiber Optic Serial reserved for connection to
one FL/PL 7000 series probe
Fiber Optic Connectors:
(2) E2000 compact duplex female for probe com-
munication and safety fiber loop
(2) FSMA for fiber connection to FM7004A
Laser: 500 mW, 830 nm
Laser shutdown time:
<1 msec after fiber disconnect
< 250 msec after loss of communication
Power Requirements:
Input Voltage 90-260 VAC, 50-60 Hz
Input Current 0.2 - 0.6 A
Input type IEC inlet with filter
Enclosure: Desktop case, 2U high; 1U blank pan-
el installed.
Dimensions (WxHxD):
without enclosure 48.3 x 4.5 x 25.4 cm (19.0 x
1.75 x 10.0 in)
with enclosure 49.8 x 12.7 x 30.5 cm (19.6 x 5.0
x 12.0 in)
Weight:
without enclosure 2.5 kg (6.25 lb)
with enclosure 7.25 kg (16 lb)
Operating Temperature Range: 10°C to 40°C
(50°F to 104°F) @ 5% to 95% RH non-condensing
Model Description
FI7000 Housed in a 2U instrument case
FI7000M1 Rack mount (Instrument case removed)
Specifications