PL7004 Kit
24 Rev F
Command To Probe Kit
Response From Probe Kit
Read Probe Axis Data (X,
Y, & Z only)
A<CR>
Probe kit response structure: <start character><command letter><x
axis field value ><y axis field value>< z axis field value><status
flag><termination>
:A<x><y><z><s><termination>
Where x, y, z, are 4 digit axis values, plus one floating decimal point
after the 2
nd or 3rd digit (in implied units of Volts/meter), and s is a
status flag (S or X ), where S indicates status is OK, X indicates status
is not OK. (See NOTE 1 on previous page.)
Described in more detail:
: = Start character ASCII colon (1 character)
A = Command identifier (1 character)
x = X axis field 4 digit value with ASCII decimal point (5 characters)
y = Y axis field 4 digit value with ASCII decimal point (5 characters)
z = Z axis field 4 digit value with ASCII decimal point (5 characters)
s = Status flag (S or X) (1 character)
termination = Depends on the selected response termination
characters (see TERM command for details) (1 or 2 characters)
Character field separators are not used.
Read Probe Identification
I<CR>
Probe kit response structure: <start character>,<command
letter>,<model>,<serial number>,<firmware rev level>,<probe
calibration date>,<status flag>,<termination>
:I,<m>,<n>,<r>,<d>,<s>,<termination>
where:
, = Separator character ASCII comma (6 characters used as separator)
: = Start character ASCII colon (1 character)
I = Command identifier (1 character)
m = probe model designation (6 characters)
n = probe serial number (8 characters)
r = probe firmware revision (10 characters)
d = probe linearization date (8 characters)
s = Status flag (S or X where S indicates status OK and X indicates
status not OK) (1 character)
termination = Depends on the selected response termination
characters (see TERM command for details) (1 or 2 characters)
Total number of characters is variable. [subject to change]