EasyManua.ls Logo

ARAG DigiWOLF 4628405 - Operation and Use

ARAG DigiWOLF 4628405
24 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16
5 USE
Once the setup is nished, Digiwolf is ready for use. The main screen shows the display divided into three horizontal sectors.
The sectors showing the symbols
1
and
2
represent the partial totalizers, individually resettable.
The third data represents instant rate value.
0.0 l
2
1
0.0 l/min
6.23 l
2
1
21.84 l/min
6.23
Feed some liquid to the system. The display will start to show the increasing value
of the measured quantity of liquid, and the instant rate value.
5.1 Partial totalizer reset
6.23 l
2
1
21.84 l/min
6.23
To reset a totalizer, it is necessary to visualize it in full.
To do this, press the button several times until the value to zero is in the central
area of the display.
6.23 l
2
1
21.84 l/min
0.00
Press ESC for two seconds. The totalizer will zero.
- Do not place DigiWolf under pressurized water.
- Use the owmeter only within recommended rate range (Tab. 2 - Par. 2.3.1). Beyond these limits, the ow-meter may give out incorrect data.
- To avoid damaging the owmeter, do not exceed the maximum rate except for short periods.
ARAG can not be held responsible for damage caused to persons, animals or things from the incorrect or unintended use of Digiwolf or
its parts.
- Do not exceed the max. operation pressure (Tab. 2 - Par. 2.3.1).
Display of the symbol ------ shows that the rate or the totalizer are over the max. displayable value.
Totalizers have oating points and display a max. of 5 gures. Up to 999.99 two decimals are shown, it then drops to 1 and 0 with the tran-
sition to thousands and tens of thousands (Fig. 43).
Partial totalizers
Instantaneous rate
Fig. 39
Fig. 40
Fig. 41
Fig. 42
Fig. 43
SETUP

Related product manuals