• Close the lid, and the instrument
weighs the weights.
• Immediately after the instrument
weighs the weights, gently open the
lid, carefully remove the 3-gram
weight, and gently close the lid.
• The test finishes once the ending
criteria has been met.
• Open the lid and remove the weight.
• Place a new/clean sample pan on the
pan support to get ready for the next
test.
• Close the lid when the heat chamber
is cool.
If desired, perform the weight test multiple times to determine the instrument’s
repeatability; this is a more precise verification of the instrument’s performance. Perform
five tests and then compute the mean, standard deviation, and relative standard deviation
(For data analysis instructions, see section 6.2 Test Results on page 21).
If all of the indicated setup steps have been completed, the instrument is now ready to
analyze samples.