Preparations
4
Actions Before First Use
1) Visually inspect the Maxi Air system for
damage.
2) Check that all parts are supplied. Compare
with section Parts Designation in this IFU. If
any part is missing or damaged - do NOT
use the product!
3) Read the IFU.
4) Disinfect the Maxi Air air supply. See
Disinfection Instructions section.
5) Perform a functionality test. See Care and
Preventive Maintenance section.
6) Find a good ventilated dry area for storing
the Maxi Air.
7) Choose a designated area where the IFU
should be kept, accessible at all times.
8) Make sure to have a rescue plan ready in
case of a patient emergency.
Below is the direction of the Maxi Air
mattress:
Actions Before Every New Patient
1) Select a suitable Maxi Air mattress size
according to sections “Correct Placement of
the Patient” and ³3URGXFW6SHFL¿FDWLRQV´.
Check that all parts of the Maxi Air is in place.
Compare with the section Parts Designation in
this IFU.
2) Carry out a thorough inspection for any
damage.
3) If any part is missing or damaged - do NOT
use the product!
4) Insert the patient name and date on the
Maxi Air mattress label. See Labels on the
Maxi Air Mattress section.
WARNING
To prevent cross-contamination:
• Never assign the usage of a mattress
to more than one patient.
• Always follow the disinfection
instructions in this Instructions for Use.
Make sure the Maxi Air air supply is disinfected
according to Disinfection Instructions section.
Actions Before Every Transfer (with
same patient)
1) Check that all parts are in place. Compare
with the Parts Designation in this IFU.
2) Carry out a thorough inspection for any
damage.
3) If any part is missing or damaged - do NOT
use the product!
NOTE
If you have any questions, please contact
your local Arjo representative for support and
service. The contact information (Arjo) is
listed on the last page of this IFU.
Head end
Foot end
RightLeft