44
FREQUENCY Initially
Before
every use
Every 2
months
Every 4
months
Every
year
Every 2
years
TRANSFER
(lifting and lowering)
0
Before
every use
500 1000 2500 5000
Load test with the SWL (maxi-
mum working capacity)
recommended.
X
Track end stoppers inspection. X
AUTHORIZED SERVICE
RAIL SECTION
FREQUENCY Initially
Before
every use
Every 2
months
Every 4
months
Every
year
Every 2
years
TRANSFER
(lifting and lowering)
0
Before
every use
500 1000 2500 5000
Torque end stoppers between 5
to 15 lbs (7 to 20 Nm).
X X
Make sure that the bracket
locking device is not visible.
XX
Make sure rail joint are closed
and that the spring pin is
centered.
X X
Make sure the rail is straight
when it is not loaded.
XX
Make sure that SWL (safe
working load) label of the rail is
equal or larger than the SWL of
the cassette.
X X
Check that the accessories
(turntable and exchanger) are
complete and correctly
maintained.
XX
Make sure that the fixations
(ceiling brackets, wall post,
wall brackets) are not
displaced.
X X
AUTHORIZED SERVICE