Refer to the GLUCOCARD™ 01-mini Blood Glucose Monitoring
System user instruction manual for further details.
Obtain a blood drop. Apply blood to the
end of the test strip. Capillary action will
pull the blood into the strip. The meter
will beep when it detects that blood has
been applied. The meter will begin to
count down. Do not add blood on top
of the test strip, as you may receive an
inaccurate result.
Important: Ensure that enough blood is applied to ll the check
window and it turns red. When the check window is lled with
blood and a beep sounds, remove the tip of the test strip from your
nger. If there is not enough blood, “E-7” will be displayed.
GLUCOCARD™ 01-mini Quick Reference Guide
Step 4
Step 5
TROUBLESHOOTING
Step 1
Step 2
Step 3
If it takes longer than
20 seconds to obtain a
blood sample and touch
the test strip to the blood
drop, repeat step 3.
PERFORMING A BLOOD GLUCOSE TEST
Quick
Reference
Guide
P/N 7347-00
Rev 7/08
Battery Symbol
Battery power is too low for testing. Replace
batteries immediately. Results may not be
accurate when testing with low battery symbol on.
Thermometer Symbol
The meter temperature is outside
acceptable range of 50˚F-104˚F (10˚C-40˚C).
Allow the meter and test strips to warm up or
cool down slowly (20 minutes) until the
temperature warning symbol goes away.
Error Message E-2
The meter temperature is outside
acceptable range of 50˚F-104˚F
(10˚C-40˚C). Allow the meter and test
strips to warm up or cool down slowly
(20 minutes) until the temperature
warning symbol goes away.
Error Message E-6
Test strip problem. Contact bars on test
strip may be dirty. Incorrect test strip
used. Used test strip has been inserted.
Repeat test with new test strip.
Error Message E-7
Additional blood was applied to test
strip after test started. Test strip was
moved during test. Insufcient blood
sample.
TEL 800.566.8558
FAX 952.646.311 0
www.arkrayusa.com
DISTRIBUTOR
Remove GLUCOCARD™ 01 SENSOR
blood glucose test strip from the bottle.
Immediately replace the bottle cap
tightly. Insert test strip as shown in
diagram. A beep will sound and the
meter will turn on automatically.
The last test result will appear. The Apply
Blood symbol and the code number
(F-2, F-3, or F-4) will alternate on the
screen. Verify that the code number on
the screen matches the code number
printed on the bottle. If the code number
does not match, the meter may
give a false reading. If code does
not match, use a new strip. If code
still does not match, use a new
bottle of test strips or call Customer
Service at 800.566.8558 (24 hours a
day, 7 days a week, USA and
Canada only).
Code No.
(F-2, 3 or 4)
Expiration date
F- 4
9
0018
F-4
Apply Blood Symbol
Code Number
When the test is done, pull out test
strip. Dispose of test strip. The meter shuts
off automatically when you remove the
test strip.
Remove lancet from lancing device.
Dispose used lancet according to
local guidelines.
The meter will count down, starting from “7”. After 7 seconds,
a beep sounds and the test result and date/time are shown.
Results will be shown in units of mg/dL.
Record result in your logbook.
83-02664
Manual
84-02681
Logbook
83-02667
Quick Reference Guide-ENG
83-02693
Quick Reference Guide-SPA
84-02690
Instruction card
84-02680
Warranty card