3.
Meter
Set-Up
4.
Trouble
shooting
Your blood glucose level is more than 600 mg/dL. Repeat test
with new test strip. If this message shows again, contact your
healthcare professional immediately.
Battery power is too low for testing. Replace batteries
immediately. Results may not be accurate when testing with low
battery symbol on.
Thermometer symbol. The meter temperature is outside
acceptable range of 50˚F-104˚F (10˚C-40˚C). Allow the meter and
test strips to warm up or cool down slowly (20 minutes) until the
temperature warning symbol goes away.
The meter temperature is outside acceptable range of 50˚F-104˚F
(10˚C-40˚C). Allow the meter and test strips to warm up or cool
down slowly (20 minutes) until the temperature warning symbol
goes away.
Test strip problem. Contact bars on test strip may be dirty.
Incorrect test strip used. Used test strip has been inserted. Repeat
test with new test strip.
Not enough blood added to test strip. Or incorrect blood
application technique. Retest with new test strip.
Code Number. This shows the code number currently in the meter.
The number on the screen must match the number on the test
strip bottle. If screen code does not match the test strip code, use
a new test strip. If code still does not match, use a new bottle of
test strips or call Customer Service at 800.566.8558.
TEL 800.566.8558
FAX 952.646.3110
www.arkrayusa.com
To enter set-up mode, hold both the
Orange/Up Arrow and Gray/Down
Arrow button for 5 seconds. “ ”
appears on the screen. Press Gray/
Down arrow button to reach desired
setting. Press Orange/Up Arrow to enter
selection and move to next setting.
<Power OFF>
Gray Arrow
Setting
the
Audible
Beep
1
Ending
Set-up
9
Setting
the Year
2
Setting the
Data
Download
Baud Rate
8
Orange
Arrow
Orange
Arrow
Orange
Arrow
Orange
Arrow
Orange
Arrow
Orange
Arrow
Return
Setting the
Hour
5
Setting the
Day
4
Setting
the
Minute
6
Setting the
Month
3
Setting
Test
Averaging
7
Orange
Arrow
Orange
Arrow
Orange
Arrow
Orange
Arrow
Your blood glucose level is less than 20 mg/dL or there was not
enough blood on test strip. Repeat test with new test strip. Make
sure you use enough blood to ll the strip. If this message shows
again, contact your healthcare professional immediately!
Meter
Components
Audible Beep
Off
Unit of
Measurement
Time Format
Control
Solution
Delete Symbol
Temperature
Warning
Date
Apply Blood
Symbol
Battery Symbol
Memory
Mode Symbol
Test Strip
Port
Gray Down
Arrow and
Orange Up
Arrow Buttons
Test Strip