Exclusive distributor of Rx100
27
CONFIDENTIAL MATERIAL
DO NOT DUPLICATE
Problem
Cause
Action
Sensation in pad(s) is
intermittent.
Examine Duty Cycle control and make
sure Duty Cycle is not set to cycle on
and off.
Set Duty Cycle to CONT (continuous).
Patient discomfort after
switching Polarity.
Output Power needs to be reduced
somewhat prior to switching Polarity to
reduce effect.
Turn Output Power down, switch
Polarity, and return power setting to a
level tolerable by patient.
Patient discomfort after
changing Main Pulse
Frequency
Output Power needs to be reduced
somewhat prior to switching changing
Main Pulse Frequency.
Turn Output Power down, switch Main
Pulse Frequency, and return power
setting to a level tolerable by patient.
The machine wont power
on.
The battery is discharged.
Charge the battery or use the external
power supply (AC adapter).
The external power supply (AC
adapter) is not plugged in.
Check the connection at the wall outlet.
Poor connection at the input power
jack.
Reposition the power plug.
The machine powers off
(the touch screen goes
blank) during a session.
The battery is discharged.
Charge the battery or use the external
power supply (AC adapter).
Poor or intermittent connection at the
input power jack.
Reposition the power plug.
Main pulses turn on
immediately instead of
gradually.
Sub-second mode has been activated.
Follow the instructions on page 15 to
deactivate Sub-second mode.