950-0073-00, Rev. B English
950-0073-00 B SM, English, Arthrex Synergy UHD4 System 1 of 74
Synergy
UHD4
System
Service Manual
AR-3200-0020
AR-3200-0021
AR-3210-0018
AR-3210-0021
AR-3210-0022
AR-3210-0023
AR-3210-0025
AR-3210-0026
AR-3210-0028
AR-3210-0029
AR-3210-0030
The Arthrex Synergy
UHD4
System Camera Control Unit and
Camera Head Service Manual provides important information
for the safe operation and service of all components of the
Synergy
UHD4
Camera System, including accessories. Read
this Service Manual thoroughly prior to using this system and
keep it in an easily accessible place for use by all operating
personnel. Read and follow all safety warnings, cautions and
precautions.
Arthrex, Inc.
1370 Creekside Blvd.
Naples, FL 34108, USA
+1 (800) 934-4404
Technical Support
1-888-420-9393
EC
REP
Arthrex GmbH
Erwin-Hielscher-Strasse 9
81249 München, Germany
+49 89 909005-0

Table of Contents

Other manuals for Arthrex AR-3200-0021

Questions and Answers

  • J
    Jonathan WebbAug 14, 2025
    What to do if there is no video image on the Arthrex Control Unit monitor?
    • P
      paige68Aug 14, 2025
      If there's no video image on the monitor, ensure the Control Unit and monitor are plugged in and powered on. Confirm all cable connections are correct, rerouting video cables if needed, and check for any cable damage, replacing them if necessary. Also, ensure the camera head cable connector is fully inserted into the Control Unit's camera head receptacle. If the issue persists, try replacing the camera head with a working one to see if the original camera head or cable is faulty and needs repair.
  • N
    Natalie WaltonAug 17, 2025
    How to fix intermittent picture on Arthrex AR-3200-0021?
    • L
      Luis RuizAug 17, 2025
      To fix an intermittent picture, first, reinsert the camera head connector card edge fully into the CCU camera receptacle. Flex the video and power cables; if the picture is affected, inspect and replace the cables as needed. If flexing the camera cable affects the picture, it may need factory repair or replacement.
  • J
    Janet OrtizAug 21, 2025
    What to do if Arthrex Control Unit camera does not power up?
    • S
      Sarah RoyAug 21, 2025
      If the camera doesn't power up and the standby LED doesn't illuminate, ensure the power cord is plugged into both the CCU and a properly grounded receptacle. If it is, try replacing the power cord, as it may be faulty.
  • D
    David BrownAug 25, 2025
    Why Arthrex Control Unit camera will not white balance?
    • J
      Jamie EvansAug 25, 2025
      If the monitor indicates “White Balance Fail”, it could be due to too much light, too little light, or the wrong color temperature light. Try moving the scope further away or closer to the white gauze during white balance, or adjust the LED Light Engine brightness accordingly.
  • M
    Melissa AdamsAug 27, 2025
    What to do if Arthrex AR-3200-0021 Control Unit tablet continually restarts?
    • A
      Alexa MillerAug 27, 2025
      If the tablet continually restarts or won't progress beyond the Samsung startup screen, first, check the tablet cable and connector for any damage. Also, ensure you're using a Samsung model EP-TA10JWE charger with a minimum of 2.0A charge current and a USB 3.0 charger cable (both connectors) no longer than 6 ft. (1.83 meters). Replace the charger or cable if they don't meet these specifications.
  • M
    monica70Aug 30, 2025
    How to fix poor color reproduction on Arthrex AR-3200-0021?
    • C
      Chad CarlsonAug 31, 2025
      To correct poor color reproduction, white balance the camera head.
  • C
    Curtis WheelerSep 3, 2025
    How to fix Arthrex AR-3200-0021 Control Unit system that does not boot up correctly?
    • B
      Bryce AdkinsSep 3, 2025
      If the system doesn't boot up correctly or banners don't display properly, power down the device, wait three minutes, and then restart the system.
  • J
    Joshua MunozSep 6, 2025
    Why Arthrex AR-3200-0021 camera head buttons do not function?
    • M
      Michael SmithSep 6, 2025
      If the camera head buttons do not function as programmed, reprogram the camera head buttons. Note that this solution applies only to the Tablet Data Input device option.
  • C
    Cory RodriguezSep 9, 2025
    What to do if AR-3200-1013 Tablet will not connect to Arthrex CCU?
    • G
      Grace MorganSep 10, 2025
      If the AR-3200-1013 Tablet will not connect to the CCU, update the SynergyUHD4 Software to v4.13.1 or newer.

Summary

Introduction and System Overview

Warnings and Precautions

Covers critical safety information, warnings, and precautions for system operation.

Critical Safety Warnings

Critical safety warnings for medical professionals to prevent severe injury or equipment damage.

Essential Precautions

Essential precautions for safe operation, user qualifications, and equipment compatibility.

Initial Device Use and Handling

System Indicators and Components

System Installation and Operation

System Installation Procedure

Provides steps for installing the Synergy UHD4 system and its components.

Typical System Installation Diagram

Illustrates the typical physical connections for setting up the Synergy UHD4 system.

System Setup Facility and Surgeon Settings

System Maintenance Configuration

Details accessing system setup through the maintenance icon and advanced settings.

System Settings Screen

Describes the 'System' screen for configuring facility name, language, and storage.

Network Configuration

Covers network configuration for connecting the system to a facility network.

Surgeon Management

Describes adding surgeons and their preferences to the system.

Surgeon Procedure Settings

Explains how to configure camera, pump, and shaver settings for specific procedures.

Scheduling and Starting Cases

System Operation Without Tablet

Camera Control Unit Components

Camera Control Unit (CCU) Description

Describes the Camera Control Unit (CCU), its components, and overall circuitry.

Camera Head Description

Describes the camera head's function, components, and connection to the CCU.

Maintenance and Cleaning Procedures

Cleaning and Sterilizing System Components

Provides instructions for cleaning and sterilizing the system components.

Camera Head Cleaning Procedures

Provides detailed steps for manual and automated cleaning of the camera head.

Sterilization of Camera Heads

Outlines steam sterilization parameters for the AR-3210 camera heads.

Troubleshooting Common Issues

Annual CCU Maintenance and Battery Replacement

Recommended Annual CCU Maintenance Tests

Specifies annual maintenance tests and their acceptable values for the CCU.

Troubleshooting Tests and Procedures

Appendix: Detailed EMC Information

Arthrex AR-3200-0021 Specifications