950-0047-04, Rev. A English
950-0047-04A Page 1 of 69
Synergy
UHD4
System
Instructions for Use Manual
AR-3200-0020
AR-3200-0021
AR-3200-0022
AR-3200-0023
AR-3210-0018
AR-3210-0021
AR-3210-0022
AR-3210-0023
AR-3210-0025
AR-3210-0026
AR-3210-0028
AR-3210-0029
AR-3210-0030
AR-3210-0031
AR-3210-0032
The Arthrex Synergy
UHD4
System Camera Control Unit and
Camera Head User’s Guide provides important information
for the safe operation of all components of the Synergy
UHD4
Camera System, including accessories. Read this User’s
Guide thoroughly prior to using this system and keep it in
an easily accessible place for use by all operating
personnel. Read and follow all safety warnings, cautions
and precautions.
Arthrex, Inc.
1370 Creekside Blvd.
Naples, FL 34108, USA
+1 (800) 934-4404
Technical Support
1-(800) 391-8599
EC REP
Arthrex GmbH
Erwin-Hielscher-Strasse 9
81249 München, Germany
+49 89 909005-0

Questions and Answers

  • M
    Madison BennettAug 14, 2025
    What to do if there is no video image on the Arthrex AR-3210-0031 Control Unit monitor?
    • B
      Benjamin StantonAug 14, 2025
      If there's no video image on the monitor, ensure the Control Unit and monitor are plugged in and powered on. Confirm all cable connections are correct, rerouting video cables if needed, and check for any cable damage, replacing them if necessary. Also, ensure the camera head cable connector is fully inserted into the Control Unit's camera head receptacle. If the issue persists, try replacing the camera head with a working one to see if the original camera head or cable is faulty and needs repair.
  • D
    Douglas ClarkAug 18, 2025
    How to fix intermittent picture on Arthrex AR-3210-0031?
    • C
      Cynthia RosarioAug 18, 2025
      To fix an intermittent picture, first, reinsert the camera head connector card edge fully into the CCU camera receptacle. Flex the video and power cables; if the picture is affected, inspect and replace the cables as needed. If flexing the camera cable affects the picture, it may need factory repair or replacement.
  • R
    ramosaliciaAug 21, 2025
    What to do if Arthrex Control Unit camera does not power up?
    • M
      Michael RiveraAug 21, 2025
      If the camera doesn't power up and the standby LED doesn't illuminate, ensure the power cord is plugged into both the CCU and a properly grounded receptacle. If it is, try replacing the power cord, as it may be faulty.
  • T
    Thomas BanksAug 25, 2025
    Why Arthrex AR-3210-0031 Control Unit camera will not white balance?
    • P
      Phyllis CarpenterAug 25, 2025
      If the monitor indicates “White Balance Fail”, it could be due to too much light, too little light, or the wrong color temperature light. Try moving the scope further away or closer to the white gauze during white balance, or adjust the LED Light Engine brightness accordingly.
  • N
    nhamptonAug 28, 2025
    What to do if Arthrex Control Unit tablet continually restarts?
    • H
      Howard HamiltonAug 28, 2025
      If the tablet continually restarts or won't progress beyond the Samsung startup screen, first, check the tablet cable and connector for any damage. Also, ensure you're using a Samsung model EP-TA10JWE charger with a minimum of 2.0A charge current and a USB 3.0 charger cable (both connectors) no longer than 6 ft. (1.83 meters). Replace the charger or cable if they don't meet these specifications.
  • A
    Amanda FlynnAug 31, 2025
    How to fix poor color reproduction on Arthrex AR-3210-0031?
    • M
      Michael CaldwellAug 31, 2025
      To correct poor color reproduction, white balance the camera head.
  • O
    omilesSep 3, 2025
    How to fix Arthrex AR-3210-0031 system that does not boot up correctly?
    • M
      michael08Sep 3, 2025
      If the system doesn't boot up correctly or banners don't display properly, power down the device, wait three minutes, and then restart the system.
  • Z
    Zachary DavisSep 6, 2025
    Why Arthrex AR-3210-0031 camera head buttons do not function?
    • X
      xspencerSep 6, 2025
      If the camera head buttons do not function as programmed, reprogram the camera head buttons. Note that this solution applies only to the Tablet Data Input device option.
  • T
    Thomas ClarkSep 10, 2025
    What to do if AR-3200-1013 Tablet will not connect to Arthrex CCU?
    • V
      Veronica NormanSep 10, 2025
      If the AR-3200-1013 Tablet will not connect to the CCU, update the SynergyUHD4 Software to v4.13.1 or newer.

Summary

SynergyUHD4 System Introduction

SynergyUHD4 Warnings and Precautions

Key terms WARNING, PRECAUTION, and NOTE explained for safe operation.

SynergyUHD4 Safety Precautions

Important safety and operational guidelines for device use and maintenance.

SynergyUHD4 Symbol Definitions

SynergyUHD4 Initial Device Use

SynergyUHD4 System Indicators

SynergyUHD4 Firmware Compatibility

SynergyUHD4 System Installation and Operation

SynergyUHD4 System Installation Steps

Step-by-step guide for setting up the SynergyUHD4 system.

SynergyUHD4 Typical System Setup

Procedure for connecting components and powering up the SynergyUHD4 system.

SynergyUHD4 Setup and Surgeon Settings

SynergyUHD4 Surgeon Procedure Settings

Configuring camera, pump, and shaver settings for specific procedures.

SynergyUHD4 Case Scheduling and Starting

SynergyUHD4 Schedule New Case

Using the "+ Add New Case" icon to schedule a new surgical case.

SynergyUHD4 Enter Case Details

Inputting patient, surgeon, and procedure information for a new case.

SynergyUHD4 Start Case

Selecting a case from the list and pressing the "Start" icon to begin.

SynergyUHD4 Case Confirmation

Verifying patient and case demographics before starting a case.

SynergyUHD4 Ending a Case

Procedure for concluding a surgical case and transitioning to review screen.

SynergyUHD4 Status Notification Icons

SynergyUHD4 Operation Without Tablet

SynergyUHD4 Initial Screen and White Balance

Initial screen display and importance of performing a white balance operation.

SynergyUHD4 White Balance Status

Visual feedback for successful (green check) or failed (red X) white balance.

SynergyUHD4 System Maintenance

SynergyUHD4 Camera Head Cleaning

Cautions and preparation for cleaning camera heads, including automated cleaning.

SynergyUHD4 Camera Head Sterilization

Steam sterilization parameters for camera heads (pre-vacuum, gravity).

SynergyUHD4 Troubleshooting Guide

SynergyUHD4 Troubleshooting Power Issues

Troubleshooting steps for when the CCU standby LED does not illuminate.

SynergyUHD4 Troubleshooting Intermittent Picture

Diagnosing and resolving issues causing an intermittent video picture.

SynergyUHD4 Troubleshooting White Balance

Troubleshooting steps for when the camera fails to white balance.

SynergyUHD4 Troubleshooting Button Function

Resolving issues where camera head buttons are not functioning correctly.

SynergyUHD4 Troubleshooting No Video Image

Troubleshooting steps for when no video image appears on the monitor.

SynergyUHD4 Error Message Resolution

SynergyUHD4 CCU Annual Maintenance

Arthrex AR-3210-0031 Specifications

General IconGeneral
BrandArthrex
ModelAR-3210-0031
CategoryControl Unit
LanguageEnglish