950-0073-00 B SM, English, Arthrex Synergy UHD4 System 71 of 74
IEC 60601-1-2
Recommended separation distances between portable and mobile RF communications equipment and the
Arthrex
UHD4
Video System
The Arthrex
UHD4
Video System is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Arthrex
UHD4
Video System can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment [transmitters] and the
Arthrex
UHD4
Video System as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum
output power of
transmitter [W]
Separation distance according to frequency of transmitter [m]
150 kHz to 80 MHz
80 MHz to 800 MHz
PEd 1/5.3
800 MHz to 2.5 GHz
PEd 1/7
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.14
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance [d] in meters [m]
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output rating of
the transmitter in watts [W] according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
PVd 1/5.3