TEST CONNECTION
Adapter Box Connections
To increase operator safety, you may elect to use an adapter box for products
terminating in either a two-prong or three-prong line cord.
Plug the black lead from the adapter box (P/N 36544) into the CONT.
CHECK terminal located on the front panel.
Plug the white lead (P/N 4040A-08) from the adapter box into the high
voltage output terminal located on the front panel.
Connect the black ground return lead (P/N 2100A-13) to the front panel
return terminal and connect the other end of the lead to the dead metal
on the chassis of the DUT. Check to ensure a solid connection is made
between the DUT and the return clip.
Plug the line cord of the DUT into the adapter box receptacle.
5
WARNING: DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST.
Note: Be sure to enable the Ground Continuity check before moving forward (see page 3).
CAUTION
HIGH VOLTAGE
5KVAC MAX.
6KVDC MAX.
NL
POWER
CAUTION
5KVAC MAX.
6KVDC MAX.
H.V.
RETURNTESTRESET CONT.
CHECK
DIELECTRIC WITHSTAND TESTER